Introduction
The continuing drive to extend the capability of optical
lithography to increasingly smaller dimensions has led to the
development of exposure tools of higher numerical aperture (NA) and
shorter wavelengths, to the exploration of new mask technologies such
as phase-shift methods [1,2], and to the refinement of mask
layouts to include optical proximity effects.
Mask design has become increasingly complex, since it is difficult to
predict the printing characteristics of a mask exposure tool
combination solely from knowledge of the layout geometry. Extensive use
of computer programs such as Splat or Image is commonly the primary
means for understanding the through-focus behavior for mask features
such as parallel lines with phase shift or vias with attenuating phase
shift. After mask design and fabrication, the printing characteristics
of the mask are studied by printing a matrix covering a range of
exposures and focus settings. These experimental test chips are then
parted, and key features are measured using a scanning electron
microscope; this is a time-consuming procedure.
The Aerial Image Measurement System (AIMS*)¹ was designed
to provide a means for rapidly evaluating the exposure and
depth-of-focus characteristics of real masks, including chrome,
proximity-corrected, phase-shifted, or attenuating phase types, prior
to resist validation. This device has been found to provide a new means
for examining masks capable of providing useful information in a very
short time before resist experiments are undertaken [3-5].
A UV microscope was set up in which the NA of the imaging lens was
adjusted so that the image produced possesses the same resolution
characteristics as that of a particular stepper. In order to emulate
effects introduced by phase shifts or optical proximity, the masks are
imaged at the wavelength of use (in the present case, 365 or 248
nm). An aperture was also placed in the illumination system so that
coherence could be matched and explored. In addition, multiple off-axis
apertures may be used to emulate off-axis illumination such as the
Canon or Nikon approach. Initial experiments were carried out in which
images were compared with extensive computer simulation and with resist
features for a wide variety of cases. Good agreement was found between
computer simulations and AIMS images obtained for test masks under a
variety of conditions. In most cases it is sufficient to predict resist
behavior via a high-contrast resist model. If desired, the image data
may be used as a starting point for more extensive resist calculations.
Tool development and description
Early in the development of phase-shift masks at IBM, it became
clear that a method of evaluating photomask performance more rapidly
than time-consuming traditional means was urgently needed. A novel
laboratory setup consisting of an industrial microscope modified to
control illumination
and objective lens NA was developed to emulate
the imaging characteristics of an optical stepper. From the prototype,
a commercial version was co-developed with Carl Zeiss Inc. and is
available as the Microlithography Simulation Microscope, MSM100
(Figure 1). Figure
2 shows a schematic layout of the MSM100 tool,
which is based upon a deep-UV Axiotron microscope with several
important differences that allow it to emulate optical steppers.
Illumination is provided by a 100-W mercury arc lamp for I-line and
deep-UV (365-and 248-nm) imaging, and by a halogen lamp for visual
inspection and alignment. Light from the mercury lamp is prefiltered by
a cold mirror, dumping excess IR radiation into a heat sink. A
narrow-bandpass filter, mounted in an automatic filter changer,
establishes the center wavelength (either 365 or 248 nm) with a
bandwidth of typically <10 nm FWHM. The coherence or
of the
light incident upon the photomask is controlled by an aperture
positioned at a point in the base of the microscope conjugate with the
objective lens pupil. This
aperture is mounted on a slider so that
various sizes may be easily selected. New optical steppers from Canon
or Nikon incorporate off-axis illumination sources to improve the
resolution of the stepper. These illumination sources can also be
studied with the AIMS microscope by inserting the appropriately shaped
aperture into the
slider. The condenser lens focuses the
illumination onto a small (submillimeter) region of the photomask. This
focusing differs from that of the optical stepper, in which a
large area of the mask is illuminated. Five- or six-inch photomasks may
be mounted on the 8 × 8-in. motorized stage. The stage is driven
vertically to collect through-focus image data. One feature of
the optical system is that a large focus motion of the microscope
stage corresponds to a small amount of defocus within a lithographic
exposure tool; thus, defocus conditions of only a fraction of a micron
may be readily simulated. The imaging system NA is controlled by an
aperture mounted on a slider in the upper column of the microscope. A
two-stage optical system magnifies the mask image onto a Photometrics
high-sensitivity UV CCD camera, which is liquid-cooled to -40°C to
ensure low noise. The 1317 × 1035-pixel-array camera has an
intensity resolution of 12 bits per pixel, sufficient for quantitative
image data analysis. The performance of the AIMS microscope was
verified for a series of s
and NAs by comparing image data with
computer simulations and resist features; good correlation was shown
[3].
Figure 1
Figure 2
Software package
An extensive software package was developed to provide a
user-friendly graphical interface for the AIMS setup, control, data
collection, and analysis. The program was written in Microsoft Visual
C++** for the Microsoft Windows® operating system. The C++ class
structure provides a modular environment for easy program
extendibility.
The program software communicates with the AIMS tool through a serial
communication port to the Zeiss MCU26 microscope stage, as well as
through a custom ISA bus adapter to the Photometrics camera. The serial
link controls the stage position, filter changer, and lamp selector.
The AIMS program, communicating with the Photometrics camera adapter,
quickly transfers image data from the camera to the computer memory.
The program provides easy management of a variety of tool control
functions. The camera type, speed, and resolution, the MCU26 controller
parameters, and the objective lens calibration file are initialized on
start-up of the program to their last saved values. All of these
parameters can be changed through program menus. An interactive
graphical display shows the current position of the lamp selector, the
filter changer, the lens turret, and the dark- and bright-field
sliders. Additional routines provide for aperture alignment, mask
focus, and mask positioning.
After loading a mask onto the stage, it is often desirable to determine
the correlation between mask and stage coordinates. By using common
alignment marks on the mask and following a three-point alignment
process, mask rotation and magnification correction factors are
determined. This permits quick and accurate mask positioning to desired
mask or stage coordinates. A KLA inspection machine is commonly used to
check a mask for defects. This inspection results in a report or list
of potential defect sites. The list is easily imported into the AIMS
program and used to move the mask swiftly from site to site to evaluate
the printability of these defects.
To assist the data acquisition process, several image-capture
parameters and routines are available: a quick image preview and
alignment mode, image size, and exposure settings. Since features on a
photomask are usually 1-3 µm in size (for 5× masks), it is not
necessary to capture and store the entire 100 × 128-µm
region of the mask that is optically imaged onto the camera. A
subfield is often specified defining the region of interest. Once set
up, single-focus images and/or through-focus images may be acquired
and stored for further analysis. For documentation purposes, a data
header is attached to each image file to record the mask designation,
measurement conditions, and user comments.
Image analysis
A variety of image analysis and data display routines are built
into the AIMS software program. Upon capture, the image is displayed on
the screen using either a linear gray-scale, pseudo-color, or
threshold-highlight palette, thus providing a qualitative picture of
the data (Figure 3). A contour plot
may then be calculated for a series of intensity threshold values
(Figure 4). The contour plot provides a quick
approximation of how the image would print in a high-contrast resist at
specified threshold values.
Figure 3
Figure 4
To provide quantitative views of the data, plots showing profile,
linewidth versus threshold, linewidth versus defocus, and exposure
defocus are available. Profile plots (Figure
5) may be calculated for a vertical or horizontal
slice of the image. The mouse may be used to select a single bright or
dark feature in the profile plot for further analysis. The linewidth
versus defocus plot (Figure 6) is determined
by calculating the width of the selected feature for specified
intensity thresholds over the range of focus. This analysis method
provides a quick means of determining expected resist feature linewidth
variations versus defocus. Finally, the exposure defocus plot
(Figure 7) is calculated by further
specifying the permitted tolerance in feature size and plotting the
limits in exposure over the range of focus.
Figure 5
Figure 6
Figure 7
These analysis methods are extremely valuable in determining the
printability of mask defects. While conventional mask inspection tools
may determine defect size, other factors (such as edge wall shape and
defect phase) are not easily measured. The AIMS tool measures the
aerial imaging performance of the mask directly, combining effects of
mask amplitude, phase, and surface topography, a result not obtained by
other inspection methods. Further analysis of the image may be
performed by importing the data file into an advanced workstation
program, for example to perform photoresist development simulations.
Photolithography involves many factors such as the reticle, exposure
system, and photoresist process which influence the final result
obtained on the substrate. These factors typically have a complex
relationship to the final results and require numerous experiments
for optimization. The typical time required to perform an experiment
capable of assessing the performance of a lithography system can range
from several days to several months, depending on the complexity of the
experiment. The AIMS technology of rapid measurement and image analysis
offers a means of emulating the lithography system by significantly
reducing the time and cost required to evaluate lithographic
performance.
Applications of AIMS
While each component of the lithography system is important to the
overall performance, the mask or reticle technology is rapidly becoming
a critical and complex element of the process. The masks/reticles are
either binary intensity masks (BIMs), in which the circuit pattern is
defined on a quartz substrate by an opaque material (Cr/CrO), or
phase-shift masks (PSMs), in which mask materials and/or topography are
utilized to delineate phase-shifted regions which modify the wavefronts
incident on the wafer through destructive interference. This
modification can result in improved resolution, exposure latitude,
and/or depth of focus relative to BIMs. Unfortunately, improved
performance is obtained at the expense of increased complexity in the
reticle fabrication process. In addition to transmission, phase must
also be controlled.
While a variety of PSM techniques exist (attenuated, alternating, rim,
and outrigger), the attenuated and alternating techniques are the most
generally studied. In the attenuated PSM, the dark areas of the mask
typically transmit the exposure energy (5-10% relative intensity)
with a 180° phase relative to the clear areas [6]. The
alternating PSM method of Levenson et al. [1] utilizes
alternating light regions to transmit exposure energy (100% relative
intensity) with a 180° phase relative to a neighboring light region
[7]. Additive or removal techniques
may be used to create alternating
phase structure (among which etched quartz technology is the most
prevalent).
Inaccurate control of fabrication parameters for a PSM can have a
detrimental impact on lithographic performance. There may be deviations
from theory, such as transmission reduction through a phase-shifted
opening which is dependent on etch roughness as well as electromagnetic
scattering phenomena from the sidewalls of the etched opening. In
addition to transmission, phase errors will exist if the correct
material depth relative to the refractive index is not obtained. This
error may be produced by etch endpoint inaccuracy during quartz
removal in a subtractive quartz process, and film property
imperfections may be produced by variations in an attenuated film
deposition process.
Fabrication processes have been developed to address these deviations
from theory in order to optimize lithographic performance. A post-etch
treatment (also referred to as an etch-back process), in which an
isotropic wet etch moves the etched-quartz sidewalls beneath the
bordering chrome film, compensates for the strong impact of
electromagnetic scattering [8].
AIMS measurements were used to
fine-tune this process on a special feature in a 0.25-µm DRAM cell
design [5], as shown in
Figure 8. The
aerial image measurement in Figure 9
demonstrates the effect of the etch-back process on the amount of
optical scattering in the phase-shifted opening. The optical scattering
produces an intensity transmission error, as seen on the no-etch case,
by its reduced peak intensity. The peak intensity of the phase-shifted
opening clearly increased relative to the non-phase-shifted opening as
the wet-etch depth increased. Optimum lithographic performance was
identified with an approximate etch-back of 1200 Å [5].
Figure 8
Figure 9
Phase optimization for alternating and attenuated PSM processes was
also established by applying the AIMS. Alternating PSM processes have
an inherent asymmetry when the phase is not ideal. This asymmetry
provides a method of extraction which relies on a comparison of the
size of adjacent features through focus. This difference in size of
adjacent openings of opposite phase
( CD) can be expressed in terms
of a phase error,
CD = A( · defocus) + B(1 - (1 - T)),
where 
is the phase error and
T is the
intensity transmission error. The phase error
defines the slope of the
CD curve. The
A and B
parameters in Equation (1) depend on the design of the alternating
pattern and exposure system NA, wavelength, and partial coherence.
AIMS measurements were used to characterize an actual alternating
PSM process. With the AIMS tool configured with
= 248 nm,
NA = 0.5, and = 0.6,
aerial image measurements were
taken through focus for a 0.20-µm alternating line/space grating,
as shown in Figure 10. The difference in
intensity peak width for lines of opposite phase was determined
as a function of focus and plotted in Figure
11 at multiple thresholds. A value of phase was
then extracted using Equation (1). The phase was found to be 5°
from optimum, requiring an etch depth correction.
Figure 10
Figure 11
Effects of the phase error observed in the AIMS measurements were also
confirmed using stepper exposures on the Micrascan® II from SVG
Lithography Systems, Inc. (SVGL). Figures 12(a) and
12(b) respectively show the 0.20-µm
pattern in 0.6 µm of APEX-E resist at 0 and 0.25-µm defocus. The
defocused condition clearly shows resist openings of differing width,
in contrast to the 0.0 defocus condition, in which all resist openings
are of uniform width. This is consistent with the aerial image, in
which neighboring intensity peaks vary in size for the defocused
condition.
Figure 12
The phase of an attenuating PSM is particularly difficult to ascertain,
since it depends on the real and imaginary parts of the refractive
index of the phase-shifting material as well as the depth into the mask
substrate. The AIMS tool, however, provides a means of measuring
the phase through the comparison of aerial image focus
characteristics of a contact and lines [9,10]. Small contact
holes receive the greatest improvement of all feature types
from attenuating phase shifting, and are therefore the most
sensitive to any phase error. The action of a phase error is to
shift the optimal focal plane in either a positive or a negative
direction. Other structures such as the line/space grating are not as
sensitive to phase, producing a significantly smaller change to the
feature's focus characteristics. By quantifying the difference in
focus at which a maximum size is achieved for a contact and nested
lines, the AIMS tool can be used to determine attenuated PSM phase
errors. Fabrication parameters can then be optimized. Figure
13 shows the difference in focus for a 0.35-µm
contact compared to a 0.35-µm line/space pattern as a function of
phase error.
Figure 13
The ability of the AIMS tool to identify fabrication errors rapidly
makes it ideal for mask qualification, in which good reticles can be
separated from bad. This is demonstrated with the DRAM cell feature in
Figure 8. Two masks were fabricated with this pattern using the same
manufacturing process. Figure 14 shows the
CD versus defocus
characteristics of the two reticles. A
significantly larger phase error exists for the first reticle, as
exhibited by the large slope of the
CD
curve versus the near-zero
slope of the second reticle. This conclusion was further supported with
exposure-defocus analysis, which showed a 62% improvement in DOF for a
15% variation in exposure for the second reticle compared to the
first. In addition, the improved performance was confirmed using
stepper exposures on the SVGL Micrascan II.
Figure 14
Lithographic performance optimization is required once fabrication
parameters are established and mask qualification is complete. The AIMS
tool provides a rapid means of optimization for stepper parameters such
as numerical aperture and partial coherence [11]. This is
demonstrated for the 0.20-µm alternating line/space grating shown in
Figure 10. Aerial image measurements were taken
through focus with the
AIMS configured at = 248 nm,
NA = 0.5, and = 0.6 and
0.3. Figure 15 shows the intensity
distributions, with the 0.3 partial coherence clearly having improved
image contrast and through-focus performance. Quantification of the
improvement is established through an exposure-defocus analysis.
Figure 15
Summary
The Aerial Image Measurement System has been introduced for
evaluating lithographic masks; its application to the development of
advanced mask designs has been described, and details of its optical
system and its software capability have been given. It has been found
that analysis of the stepper equivalent image produced by the AIMS tool
provides a new capability for rapid evaluation of the printing
properties of chrome and phase-shifted masks prior to undertaking
extensive resist-validation exposures and SEM feature size
measurements. Examples have been presented demonstrating its
application to the optimization of the mask fabrication process,
measurement of mask phase error, characterization of feature
printability, and optimization of mask photolithographic performance.
Acknowledgments
The authors wish to thank Omesh Sahni for his continuing support
and encouragement, as well as colleagues at IBM and Carl Zeiss Inc. for
many useful interactions.
*AIMS is a trademark of International Business Machines
Corporation.
**Visual C++ is a trademark, and Microsoft Windows is a
registered trademark, of Microsoft Corporation.
Micrascan is a registered trademark of SVG Lithography Systems,
Inc.
References
¹ Patent applied for in 1994.
Received February 9, 1996; accepted for publication October 15,
1996
|